Isoplex MFM-FLOW Gas Mass Flow Meter

Isoplex MFM-FLOW series features a closed-loop control system that integrates a high-precision mass flow sensor, a laminar flow element, an electromagnetic control valve and a precision amplification circuit. At its core lies the principle of calorimetry based on the temperature difference across a capillary tube; by measuring changes in the distribution of heat conduction caused by the flow of the fluid, it enables direct detection of the mass flow rate of the gas. This series not only delivers highly accurate flow measurement but also utilises an internal closed-loop control system to automatically actuate the control valve, ensuring that the actual flow rate strictly matches the setpoint signal.

Core Operating Principle

  • Thermal Sensing: As the fluid flows through the measuring capillary, it carries away heat from the upstream side of the sensor and transfers it downstream.
  • Signal Processing: The sensor detects the temperature difference between the upstream and downstream sides of the capillary, and this signal is precisely amplified by a multi-layer circuit board.
  • Closed-Loop Control: The control circuit continuously compares the sensor signal with a preset target value in real time, and uses the difference to drive the control valve, dynamically compensating for flow fluctuations.

Technical Features & Ventajas

  • Thermal Mass Flow Principle

The measurement results directly reflect the number of gas molecules. The measurement process requires no temperature or pressure compensation, effectively eliminating interference from environmental fluctuations on control accuracy.

  • Structural Stability

The flow divider features a laminar flow stack design. Compared to traditional structures, the laminar flow stack provides a more stable flow field distribution, ensuring excellent linearity and long-term operational reliability across a wide measurement range.

  • Excellent Performance at Ultra-Low Flow Rates

Thanks to the high-sensitivity sensor design, this product features short warm-up times and minimal zero-point drift. Even during ultra-low flow control tasks, it maintains extremely high setpoint repeatability and measurement accuracy.

  • High-Immunity Electronic System

The circuit board employs a multi-layer routing process. This design not only reduces the device’s size but also significantly enhances the circuit’s electromagnetic compatibility and immunity to interference, ensuring signal purity even in complex industrial electromagnetic environments.

Application Area

Semiconductor processes, laboratory analysis, fuel cell R&D, environmental monitoring, and precision vacuum coating.

Visual Identification

To help users easily distinguish between different gases, we have designed three color schemes, which can also be customized to meet specific requirements.

Red: Toxic and hazardous gases, such as hydrogen sulfide and ammonia

Yellow: Flammable and explosive gases, such as hydrogen and oxygen

Gray: Inert gases, such as air and nitrogen

Visual Identification

To help users easily distinguish between different gases, we have designed three color schemes, which can also be customized to meet specific requirements.

Red: Toxic and hazardous gases, such as hydrogen sulfide and ammonia

Yellow: Flammable and explosive gases, such as hydrogen and oxygen

Gray: Inert gases, such as air and nitrogen

Model Selection

Isoplex MFM FLOW Gas Mass Flow Meter

Specification of Isoplex MFM-FLOW Gas Mass Flow Meter

 

Low Range

Medium Range

Large Range

 

Analog

Local Display

Local Display

Digital

Analog

Local Display

Gama

5~500 mln/min,1~30 ln/min (Air)

30~300slm (Air)

300~3000slm(Air)

Precisión

±1.0%F.S(Flow≤10 slm)

±1.5%F.S(10 slm<Flow≤30 slm)

±1.0%S.P(Flow≥35%F.S)

±0.5%F.S(Flow<35%F.S)

±1.0%S.P(Flow≥35%F.S) ±0.5%F.S(Flow<35%F.S)

±1.0%S.P(Flow≥35%F.S) ±0.5%F.S(Flow<35%F.S)

±1.0% F.S

±1.0%S.P(Flow≥35%F.S) ±0.5%F.S(Flow<35%F.S)

Control Range

Flow Controller:50:1,Flow Meter:100:1

Flow Controller:50:1

Flow Controller:50:1

Repeatability

±0.2%F.S

±0.2%F.S

±0.2%F.S

±0.2% F.S

±0.2% F.S

±0.2%F.S

Linearity

±0.5%F.S(Flow≤10 slm),±1.0%F.S(10 slm<Flow≤30 slm)

±0.5%F.S

±0.5%F.S

±0.5% F.S

±0.5% F.S

±0.5%F.S

Temperature sensitivity

zero: < ± 0.05% FS/℃

span: < ± 0.1% F.S/℃

zero: < ± 0.05% FS/℃ span: < ± 0.1% F.S/℃

zero: < ± 0.05% FS/℃ span: < ± 0.1% F.S/℃

zero: < ± 0.05% FS/℃ span: < ± 0.1% F.S/℃

zero: < ± 0.05% FS/℃  span: < ± 0.1% F.S/℃

Response Time

1~4 sec (T98)

<2 sec (T98)

1~4sec (T98)

1~4 sec (T98)

1~4 sec (T98)

1~4 sec(T98)

Environmental temperature

5~45℃

5~45℃

(Ensure Accuracy

:15~35℃)

5~45℃

(Ensure Accuracy

:15~35℃)

5~45℃

(Ensure Accuracy

:15~35℃)

Leak integrity, outboard

<1×10-8Pa·M³/S(He)

<1×10-10Pa·M3/S(He)

<1×10-10Pa.M3/S(He)

< 1×109Pa·M3/S(He)

< 1×10-8Pa·M3/S(He)

<1×10-10Pa.M3/S(He)

Warm-up Time

15 min

Power Supply

15~24 vdc

Analog Signal

0~5 vdc or 4~20 mA

0~5 vdc or 4~20 mA

Input/Output Signal

RS485/RS232

(0~5 vdc or 4~20 mA

RS485/RS232

(0~5vdc or 4~20mA)

RS485/RS232

 (0~5vdc or 4~20mA)

RS485/RS232

(0~5vdc or 4~20mA)

Primary Wetted Materials

SS316 Stainless Steel

Max Pressure Rating

3 MPa

(Special: 10 MPa)

3 MPa

30bar

30 bar

30 bar

3MPa

Max Operating Pressure

1 MPa

Pressure difference

0.05 – 0.4 MPa

Mechanical connection

Standard:Double ferrule connection

(Options:Thread VCR VCO or Flange)

Seals

Viton, NBR, EPDM, Perfluoroethane

Viton, NBR, EPDM, Perfluoroethane

Appearance of Isoplex MFM-FLOW Gas Mass Flow Meter

Isoplex MFM FLOW Gas Mass Flow Meter
Analog
Low Range
Isoplex MFM FLOW Gas Mass Flow Meter
Local Display
Low Range
Isoplex MFM FLOW Gas Mass Flow Meter
Local Display
Medium Range
Isoplex MFM FLOW Gas Mass Flow Meter
Digital
Medium Range
Isoplex MFM FLOW Gas Mass Flow Meter
Analog
Medium Range
Isoplex MFM FLOW Gas Mass Flow Meter
Local Display
Large Range

Download Documents

¿Por qué elegir Biovanix?

Conocimientos y experiencia

Con años de experiencia en la industria de transferencia de fluidos, tenemos la experiencia y el conocimiento para ofrecerle las mejores soluciones para sus necesidades específicas. Nuestro equipo de profesionales se dedica a ayudarle a elegir la bomba de émbolo buzo adecuada para su aplicación y a garantizar el éxito de su implantación.
 

Calidad y fiabilidad

Nos comprometemos a ofrecerle productos y servicios de la máxima calidad. Nuestra bomba de émbolo buzo está diseñada y fabricada con los más altos estándares, garantizando un rendimiento fiable y una larga durabilidad.
 

Atención al cliente

En [Nombre de su empresa], creemos que la satisfacción del cliente es nuestra máxima prioridad. Nuestro equipo de atención al cliente está siempre disponible para ayudarle con cualquier pregunta o problema que pueda tener. Desde el soporte técnico hasta el servicio posventa, estamos aquí para garantizarle una experiencia sin problemas ni complicaciones.
 

Soluciones a medida

Entendemos que cada aplicación es única, por lo que ofrecemos soluciones personalizadas adaptadas a sus necesidades específicas. Tanto si necesita una bomba de émbolo estándar como una solución personalizada, tenemos capacidad para satisfacer sus requisitos.

Póngase en contacto con nosotros

¿Está listo para experimentar la potencia y fiabilidad de nuestra bomba de émbolo buzo? Póngase en contacto con nosotros hoy mismo para obtener más información sobre nuestros productos y servicios. Nuestro equipo de expertos está listo para ayudarle y proporcionarle la mejor solución de transferencia de fluidos para su aplicación.
Scroll al inicio
biovanix-logo

Obtenga un presupuesto rápido